IMR EX610 User manual

R2.1
Environmental Equipment, Inc.
IMR EX610 Portable SMART Sensor Gas Detector
User Manual
Read this manual carefully before using this device.
(727)328-2818 / (800)RING-IMR • Fax: (727)328-2826
www.imrusa.com • email: inf[email protected]om

®
Environmental Equipment, Inc.
IMR EX610 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 2of 27
Table of Contents
Table of Contents.................................................................................................................................................... 2
Safety Information.................................................................................................................................................. 4
1. Brief Introduction ............................................................................................................................................ 5
2. Main Function and Specification..................................................................................................................... 5
2.1 Main Function and Features..................................................................................................................... 5
2.2 Technical Specification............................................................................................................................ 6
3. Configuration and function.............................................................................................................................. 7
3.1 Appearance............................................................................................................................................... 7
3.2 Display information.................................................................................................................................. 8
3.3 Buttons ..................................................................................................................................................... 9
4. Operation ....................................................................................................................................................... 10
4.1 Power on................................................................................................................................................. 10
4.2 Power off................................................................................................................................................ 11
4.3 Bump Test.............................................................................................................................................. 12
4.4 Settings Menu......................................................................................................................................... 12
4.5 Alarm information.................................................................................................................................. 14
4.6 Status Checking...................................................................................................................................... 15
4.7 Zero Point Adjustment........................................................................................................................... 16
4.8 Calibration and alarm adjustment .......................................................................................................... 16
4.8.1 Calibration and Alarm Levels......................................................................................................... 16
4.8.2 Warming Up.................................................................................................................................... 17
4.8.3 Zero Point Calibration..................................................................................................................... 17
4.8.4 SPAN Calibration Setup................................................................................................................. 18
4.8.5 SPAN Calibration ........................................................................................................................... 18
4.8.6 Low Alarm Level Setting (L-Alarm).............................................................................................. 19
4.8.7 High Alarm Level Setting (H-Alarm)............................................................................................. 19
5. Data Uploading (If Equipped)....................................................................................................................... 20
6. Using and Replacement of sensors................................................................................................................ 20
7. Trouble Shooting Guidance........................................................................................................................... 20

®
Environmental Equipment, Inc.
IMR EX610 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 3of 27
8. WARNINGS.................................................................................................................................................. 21
9. Chart 1 - Gas List........................................................................................................................................... 22
10. WARRANTY ............................................................................................................................................ 23
11. Contact Information................................................................................................................................... 24

®
Environmental Equipment, Inc.
IMR EX610 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 4of 27
Safety Information
Please read the following information before using the detector:
Do not use a damaged gas detector. Before using the detector. Check for cracks or missing parts. If the
detector is damaged or incomplete, please contact IMR.
It is highly recommend to do a "bump test" to ensure proper performance of the detector (refer to section
4.3 for more information). Calibrate the device if the test value is beyond detection range (refer to
section 4.8 for more information).
Do "bump tests" periodically to confirm sensor is responding to gas. Verify that the audio, visual, and
vibration alarms are working properly during the test.
Use only accessories specially designed for EX610 or accessories approved by IMR.
Use only the charger supplied with the detector. Do not charge the detector in hazardous area.
Detectors with catalytic or semiconductor sensor: Exposing the sensor to gas concentrations beyond its
detection range for a prolonged period of time can adversely affect the cell and possibly lead to damage.
Detectors with catalytic or semiconductor sensor: Will be damaged if exposed to environments
containing: lean chromate, sulfur compounds, phosphorus compounds, or silicon. Do not use the
detector in these environments.
Do not expose the device to environments that consist of hydrogen sulfide, halogenated hydrocarbon, or
highly corrosive gases for extended periods of time. Doing so will reduce the response of the sensor and
there for its accuracy. If the device needs to be used in environments preform a Bump Test before and
after to ensure functionality of the sensors (see section 4.3).
Do not expose the device to environments that have the potential for electric shock, strong magnetic
fields, or intense continuous mechanic vibration.
This device uses a lithium-ion battery. Follow all local environmental and governmental regulations
when disposing of the battery.
Any unauthorized disassembly, modification or repair of the gas detector will void the warranty.
Avoid dropping the detector from high elevations or exposing it to extreme vibrations.
For any issues not covered in this manual. Contact IMR for more information.

®
Environmental Equipment, Inc.
IMR EX610 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 5of 27
1. Brief Introduction
IMR EX610 portable single gas detector is an intrinsically safe instrument which continuously detects
combustible and toxic gases. It is suitable for combustible and toxic gas leakage detection in underground pipes
or mines. It keeps workers safe and helps prevent facilities from being damaged.
The detector uses high-quality sensors and detects gases via the natural diffusion method. It also has high
sensitivity and repeatability. The detector uses an embedded MCU controller making it easy to operate.
The shell uses special high strength material and anti-slip rubber. It is also water-proof and dust-proof.
2. Main Function and Specification
2.1 Main Function and Features
Advanced 16-digit MCU controller with low power consumption
Wide view LCD screen
Adjustable high alarm and low alarm setting
Adjustable calibrating level for easy calibration
Devices with LEL range have a high-concentration protection function
Self-tests combustible gas sensors
Low-voltage alert
Real time clock display
Smart Sensor module design for easy maintenance and replacement
Self-adjustment function to reduce detection errors
Audio, visual, and vibration alarms
Upload data to a PC
STEL and TWA alarming for toxic gases
Self-test at power on with advanced fault diagnosis and auto reset
Password management to avoid improper operation or configuration
Intrinsically safe design

®
Environmental Equipment, Inc.
IMR EX610 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 6of 27
2.2 Technical Specification
Ranges: See attached table 1.
Gases Detected: combustible gas, toxic gas, oxygen
Response time:
oSemiconductor, Catalytic, Thermal conductivity t90<30s
oOxygen, Carbon monoxide, hydrogen sulfide t90<30s
Indication method:
oLCD displays the real time data and status
oAlarm method: Audio, video and vibration alarm
Working Temperature:
oCombustible Gas: -40℃~70℃
oToxic Gas -20℃~50℃
Working humidity:≤95%RH no condensation
Power source: DC3.7V (Lithium battery 1800mAh)
Charging time: 4~6 hours
Continuous working time:
o≥8 hours (for combustible gas)
o≥300 hours (for toxic gas under temperature of 20℃)
Protection Grade: IP66
Weight: about. 191g (with battery)
Dimensions: 108.4mm×61mm×36mm (L × W × H)

®
Environmental Equipment, Inc.
IMR EX610 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 7of 27
3. Configuration and function
3.1 Appearance
#
Description
1
Alarm light
2
LCD Screen
3
Buzzer port
4
Sensor Port
5
Button
6
Buckle

®
Environmental Equipment, Inc.
IMR EX610 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 8of 27
3.2 Display information
#
Function
1
Battery Status
2
Lock on or Password
3
Buzzer Status
4
Vibration Status
5
Alarm Information
6
Warning or Max Value
7
Gas Concentration
8
Error
9
Record
10
Clock Icon
11
Zero Calibration
12
Calibration Point or Calibration Status
13
Communication icon ①
14
Data Setting
15
Engineering Unit
Note: ①only appears when the device is uploading data.

®
Environmental Equipment, Inc.
IMR EX610 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 9of 27
3.3 Buttons
Button
Function
Press and hold for more than 3 seconds to turn on the device.
Press to cancel configuration setting.
When powered off. Press and for more than 5 seconds to calibrating the device.
Hold for a few seconds to turn off the device.
Press to adjust a value
Press to enter the setup menu for the selected item
Press and for more than 3 seconds to access the configuration menu.
Press and to review temperature, time, STEL , TWA①and Max Value ②
Press to check setup
Press to save a setting.
If the alarm is active. Press to cancel an audible alarm and vibration.
Notice:①Only available for toxic gas detector
②Oxygen has Max/Min exposer value

®
Environmental Equipment, Inc.
IMR EX610 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 10 of 27
4. Operation
4.1 Power on
With the device powered off. Hold for more than 3 seconds and the device will start up. The device then
starts a self-test as follows:
1. All the data and figures displayed; back light on
2. Buzzer sounds
3. All alarm functions
4. Version number displayed
5. Date & Time display
Year Date Time
6. Gas type
7. L-alarm and H-alarm
L-alarm H-alarm
Table of contents
Other IMR Gas Detector manuals
Popular Gas Detector manuals by other brands

Macurco
Macurco GD-6 Operation manual

Evikon
Evikon E2632 user manual

Critical Environment Technologies
Critical Environment Technologies CGAS-A Series Installation & operation manual

MSA
MSA altair 5 operating manual

DOD Technologies
DOD Technologies ChemLogic CL96 operating manual

DEGA
DEGA NS III LCD Series instruction manual