LVB-2 Instruction Sheet
© Ikonix USA 3
GENERAL CONTINUED
TEST POINTS TEST CONDITION SPECIFICATION
FAIL 30kΩ 1. LLT: 132VAC, 2000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%
PASS 350kΩ 1. LLT: 132VAC, 500uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%
FAIL 150kΩ 1. LLT: 132VAC, 500uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%
PASS 1.47MΩ 1. LLT: 132VAC, 100uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%
FAIL 350kΩ 1. LLT: 132VAC, 100uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%
PASS 2.97MΩ 1. LLT: 132VAC, 30uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%
FAIL 1.47MΩ 1. LLT: 132VAC, 30uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%
PASS 30kΩ 1. LLT: 264VAC, 10,000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%
FAIL 15kΩ 1. LLT: 264VAC, 10,000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%
PASS 150kΩ 1. LLT: 264VAC, 6000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%
FAIL 30kΩ 1. LLT: 264VAC, 6000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%
PASS 350kΩ 1. LLT: 264VAC, 2000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%
FAIL 150kΩ 1. LLT: 264VAC, 2000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%
PASS 1.47MΩ 1. LLT: 264VAC, 500uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%
FAIL 350kΩ 1. LLT: 264VAC, 500uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = O
2%