outside) including the cavities of the Spray Gun. The components must be
cleaned with a non-linting wipe, sponge, or brush.
2. Place the individual components in the ultrasonic and observe the compatibility
of the ultrasonic cleaner and the contact time.
3. Rinse off the cleaning gun and the components with sufficient water and remove
any remaining detergent residue.
4. Dry the outside and inside with a non-linting wipe; use a medical compressed
air.
●It is recommended to decalcify the Spray Gun approximately when needed.
oDisassemble the gun and place the inner parts into a decalcifying agent.
oThis is beneficial for the service life and reliability of the Spray Gun.
●It is not necessary to carry out water quality tests in the case of using potable
water.
●Replacement parts and instructions for repair are available from Healthmark.
●Component Warranty:
oStainless-steel parts have a one (1)-year warranty.
oO-Rings are worn parts and do not fall under warranty.
oIn the event of improper intervention of the mechanical components of
the cleaning gun, the statutory guarantee and warranty claims expire.
Reassembly of the Spray Gun after sterilization:
The complete assembly of the cleaning gun must be carried out under sterile conditions.
Before screwing the individual parts together, make sure all O-Rings are in
perfect condition.
Sealing elements must be moistened with a suitable physiological saline solution
before use.
Sealing rings are to be mounted at the intended locations and locked in place by
pressing them firmly by hand.
1. O-Ring - Pos. (2): Place the moistened O-Ring on the attachment Pos. (1).
2. Hose attachment - Pos. (1): Screw the attachment to the gun body by turning it
clockwise.
3. Dosing trigger - Pos. (3): Attach the dosing trigger to the gun body.
4. Dosing lever axle - Pos. (4): Use your fingers to completely push the dosing
lever axle into the gun body to lock the dosing lever. An additional operation of
the dosing trigger facilitates the movement.
5. Dosing pin - Pos. (7): Insert dosing pin with moistened (O-Ring 3 x 1 mm) into
the gun body. If necessary, press the dosing pin with your finger to prevent it
from falling out. (Attention: If the dosing pin is inserted with force, the O-Ring
may be damaged and the gun may leak at the dosing pin position.)
6. Sealing cone - Pos. (8): Insert the sealing cone into the gun body with the short
side facing toward the gun.
7. Dosing spring - Pos. (7): Place the dosing spring on the long side of the sealing
cone and insert it into the gun body.
8. O-Ring - Pos. (10): Put the moistened O-Ring on the sealing cap Pos.: (1,1) and
push it into the intended position.
9. Sealing cap - Pos. (1, 1): Turn the cap clockwise to carefully screw it onto the
gun body against the pressure of the dosing spring.
10. Silicone handle - Pos. (1, 2): Put the silicone handle on the gun body. The
inscription must point to the front as shown in the illustration.
11. O-Ring - Pos. (1, 3): Put the moistened O-Ring on the gun spout Pos.: (1, 4) and
press it into the intended position.
12. Holding cone - Pos. (1, 5): Screw and tighten the holding cone clockwise with
the hose to the gun body and tighten sufficiently.