Hanatek CBT1 User manual

CBT1.2 Operating Manual Issue C
28-05-19
0
HANATEK
Crease and Board Stiffness Tester
CBT1
Model 9011
INSTRUCTION MANUAL
Rhopoint Instruments
Hanatek Instruments
Rhopoint House
Enviro 21 Park
Queensway Avenue South
St Leonards on Sea
TN38 9AG UK
Tel. No. +44 (0)1424 739623
Fax No. +44 (0)1424 730600
www.hanatekinstruments.com

CBT1.2 Operating Manual Issue C
28-05-19
1
Contents
Page
Introduction 2 - 3
Instrument menus and screen 4 - 9
Identifying grain direction 10
Sample for board testing 11
Sample for crease testing 12 - 13
Board test 14 - 16
Crease test 17
Interpretation of results 18 –19
Exporting results 20 - 21
Calibration 22
Maintenance 23

CBT1.2 Operating Manual Issue C
28-05-19
2
The Hanatek Crease and Board Stiffness Tester
1. Identifying parts on the crease and board stiffness
tester
The crease and board stiffness tester measures the force
required to hold a creased sample folded to 90oor a board
sample bent through 15o, the results being taken after these
conditions have been maintained for 15 seconds (s). The
samples are bent against the load cell blade (F) and the force on
this blade is displayed on the LCD display (C).
Operating the instrument is through rotating the relevant test jaw
(J + I) and using the capacitive touch buttons (D)
F
C
D
G
J
H
F
I
H

CBT1.2 Operating Manual Issue C
28-05-19
3
The display automatically displays the force reading at the end of
the 15s test period. Both results from crease testing and board
stiffness testing are stored in separate columns with batch testing
stats and the crease to board stiffness ratio also displayed.
To operate, the instrument needs connection to a 110-240V AC
supply via the AC-DC 24V power adapter. The power DC power
input socket and ON switch is located at the rear of the instrument
(G).
At switch-on, the display (C) will display the instrument serial
number, last certified date and the firmware version loaded into
the three PCBs inside the CBT instrument. This information may
be requested by Hanatek or approved repair centres if you
require technical assistance at any stage.
It is recommended with any measuring equipment that it is left for
thirty minutes to warm up before commencing testing. The
recommended operating conditions as stated in the standard
BS6965 is 23deg C at 50% r.h.

CBT1.2 Operating Manual Issue C
28-05-19
4
Instrument Menus and operating screens
The instrument has been designed for ease of operation with
options shown in images and symbols where ever possible.
Start up screen
On first power up Initialising hardware please wait will be written
on the screen below the firmware version numbers.
During this period do NOT touch the cap buttons as they are
calibrating.
Wait until the screen display is as below and then tough any of
the cap sense buttons to continue.
Cap buttons are highlighted with a red X
Last Certified 28/03/2014
SN: CBT01010101A
Ver: G: 2.00 T: 1.00 L: 4.00
X
X
X
X
X
X

CBT1.2 Operating Manual Issue C
28-05-19
5
Main Operating Screen
The main operating screen is shown below. On the left-hand side
are the following Icons.
Enter Options Menu
Tare/Zero the instrument
Output test data
Max:
Min:
Mean:
SD:
Samples:
Max:
Min:
Mean:
SD:
Samples:
- - - - - -

CBT1.2 Operating Manual Issue C
28-05-19
6
Options Screen
The options screen is shown below. On the left-hand side are the
following icons.
Adjust time and date
Adjust test time from 10-15 Seconds *
15seconds is specified in the standard BS6965 and the recommended time.
Turn Stats display on and off
Return to main test screen
15 s
On

CBT1.2 Operating Manual Issue C
28-05-19
7
Adjusting the time and date
From the options screen press the button next to the Icon.
Press the middle button on the bottom button bar to enable
adjustments.
Use the up and down arrows to adjust the green highlighted field.
Pressing the middle button will skip to the next field.
When finished press the left hand button to save and exit to the
main screen
Return to main test screen Set Time and Date
11 : 44 : 51
April 04
2014

CBT1.2 Operating Manual Issue C
28-05-19
8
Adjusting the test time
From the options screen press the button next to the Icon.
This will cycle the test time from 10-15 seconds.
The Standard BS6965 recommends 15seconds. It is
recommended that the operator does not deviate from this.
Turning Stats On and Off
From the options screen press the button next to the Icon.
This will cycle from On - Off.
Return to main test screen
15 s
On

CBT1.2 Operating Manual Issue C
28-05-19
9
Using the Instrument
At the beginning of testing, and at regular intervals during
extended testing, it is recommended that the instrument tare
button is pressed to re-zero.
Before placing a sample in the clamp, Press the cap sense button
next to the icon on the screen (K).
The sample for testing may now be inserted in the appropriate
clamp (J) for board stiffness, (I) for crease. To open the clamp,
rotate the knob (H). Further details of setting up samples in the
clamps are given later in the text. On turning the appropriate jaw
to fold the sample the timing sequence is automatically initiated.
After 15 s the timer the stiffness result is displayed on the display.
This figure is held until the next sequence.
T
Max:
Min:
Mean:
SD:
Samples:0
Max:
Min:
Mean:
SD:
Samples:0
14s
- - - -
Max: 44.4g
Min: 44.0g
Mean: 44.2g
SD: 0.173
Samples: 4
Max:
Min:
Mean:
SD:
Samples:0
44.1g
- - - -
Max: 44.4g
Min: 44.0g
Mean: 44.2g
SD: 0.173
Samples: 4
Max:
Min:
Mean:
SD:
Samples:0
44.1g
- - - -
When the sample jaw is rotated the
countdown timer will start and the
clock will show at the bottom of the
screen
When the countdown timer has
finished the X will show at the
bottom of the screen indicating the
jaw needs to be returned.
The bin Icon is shown allowing the
user to discard the last test result.
Pressing and holding the button will
delete all the samples.
Max: 44.4g
Min: 44.0g
Mean: 44.2g
SD: 0.173
Samples: 4
Max: 43.9g
Min: 43.7g
Mean: 43.8g
SD: 0.099
Samples:4
44.1g
0.992
Main test screen showing 4 crease
and 4 board tests. The ratio has
been calculated automatically.
43.7g
This manual suits for next models
1
Table of contents