1. INTRODUCTION
1-1. SPECIFICATIONS
SPEC MODEL OS – 5060 A
6-inch rectangular screen with internal graticule ; 8X10 Div
(1div = 1Cm), making for measurement of rise time.
2mm subdivisions along the central axis.
+10Kv approx. (ref.cathode)
P31 (standard)
Possible(with autofocus correction circuit)
provided
variable
* CRT
1) Configuration
2) Accelerationg potential
3) Phosphor
4) Focussing
5) Trace rotation
6) Scale illumination
7) Intensity control provided
Positive going signal decreases intensity +5Vp-p or more
signal cases noticeable modulation at normal intensity setting
DC-3.5MHz (-3dB)
DC
20k – 30k ohms
* Z-Axis input
(Intensity Modulation)
1) Input signal
2) Band-width
3) Coupling
4) Input impedance
5) Maximum input voltage 30V(DC+peak AC)
(X1) DC to 60MHz normal
(X5) DC to 20MHz magnified
(X1) 10Mz to 60MHz normal
(X5) 10Mz to 20MHz magnified
CH1, CH2, ADD, DUAL(CHOP: Time/div switch 0.2s~5ms,
ALT:Time/divswitch2ms~0.1㎲)
5mV/div to 5V/div in 10calibrated steps of a 1-2-5 sequence.
Continuously variable between steps at least 1 : 2.5
X5 MAG : 1 mV /div to 1V/div in 10calibrated steps.
Normal: ±3% magnified: ±5%
* Vertical Deflection(1)
1) Band-width(-3dB)
DC coupled
AC coupled
2) Modes
3) Deflection Factor
4) Accuracy
5) Input impedance Approx. 1M-ohm in parallel with 25pF
6) Maximum input voltage Direct : 250(DC+peak AC), with probe :
refer to probe specification
7) Input coupling DC – GND – AC
7