EKOM DK50 DE User manual


GENERAL INFORMATION
2 NP-DK50 DE-LF-9_02-2019
CONTENTS
1.GENERAL INFORMATION ........................................................................................... 3
PURPOSE ................................................................................................................ 3
CONTRAINDICATIONS AND SIDE-EFFECTS ........................................................ 3
OPERATOR'S RESPONSIBILITY FOR PATIENT SAFETY ..................................... 3
MARKINGS ............................................................................................................... 3
WARNINGS .............................................................................................................. 3
General safety warnings .................................................................................. 4
Electrical system safety warnings .................................................................... 4
WARNING NOTICES AND SYMBOLS ..................................................................... 5
USE .......................................................................................................................... 6
STORAGE AND TRANSPORT ................................................................................. 6
2.EQUIPMENT DESCRIPTION ........................................................................................ 7
3.TECHNICAL DATA ....................................................................................................... 9
4.OPERATION ................................................................................................................ 11
INSTALLATION AND FIRST OPERATION ............................................................ 11
Removal of transport stabilizers .................................................................... 11
Compressed air connection ........................................................................... 12
Electrical connection ...................................................................................... 12
First operation................................................................................................ 12
Adjusting the output air pressure ................................................................... 13
OPERATION ........................................................................................................... 13
Switching the compressor on ........................................................................ 13
Running the compressor ............................................................................... 13
Cleaning the compressor ............................................................................... 13
5.MAINTENANCE .......................................................................................................... 14
REPAIRS AND SERVICE ....................................................................................... 14
Cover removal ............................................................................................... 14
MAINTENANCE SCHEDULE ............................................................................... 14
Checking tightness of joints and inspecting the equipment ........................... 15
Replacing filter elements ............................................................................... 16
Changing suction filter elements.................................................................... 16
Securing the equipment before transport ...................................................... 17
SHUTDOWN ........................................................................................................... 17
EQUIPMENT DISPOSAL ........................................................................................ 17
6.TROUBLESHOOTING ................................................................................................ 18
7.SPARE PARTS ........................................................................................................... 19
8.ELECTRIC AND PNEUMATIC DIAGRAMS................................................................ 20
WIRING DIAGRAM ................................................................................................. 20
PNEUMATIC DIAGRAM ......................................................................................... 20
9.ELECTROMAGNETIC COMPATIBILITY DECARATION ........................................... 22
10.ANNEX ........................................................................................................................ 95
INSTALLATION RECORD ...................................................................................... 95

GENERAL INFORMATION
NP-DK50 DE-LF-9_02-2019 3
1. GENERAL INFORMATION
PURPOSE
The EKOM DK50 DE is a medical air compressor that supplies clean, oil-
free compressed air for use with medical ventilators.
CONTRAINDICATIONS AND SIDE-EFFECTS
There are no contraindications or side-effects known.
OPERATOR'S RESPONSIBILITY FOR PATIENT SAFETY
Instructions for use is an integral part of the equipment and must be kept
with the compressor. Careful review of this manual will provide information
necessary for correct operation of the equipment.
Rx only
US Federal law restricts the sale of this device by or on the order of a
physician.
MARKINGS
Products marked with the CE mark of compliance meet the safety
requirements of the European Union (93/42/EEC).
WARNINGS
The safety of operating personnel and trouble-free operation of the
equipment are ensured only if original parts are used. Only accessories
and spare parts mentioned in the technical documentation or expressly
approved by the manufacturer can be used.
If any other accessories or consumable materials are used, the
manufacturer cannot be held responsible for the safe operation and
functionality of the equipment.
The warranty does not cover damages resulting from the use of
accessories or consumable materials other than those recommended by
the manufacturer.
The manufacturer assumes responsibility for the safety, reliability and
function of the equipment only if:
- Installation, calibration, amendments, extensions and repairs are
performed by the manufacturer, one of its representatives or a service
provider authorized by the manufacturer
- The equipment is used in accordance with the Installation, Operation and
Maintenance Manual

GENERAL INFORMATION
4 NP-DK50 DE-LF-9_02-2019
The Installation, Operation and Maintenance Manual accurately describes
the design of the compressor and its compliance with safety and technical
standards. The manufacturer reserves all rights to its wiring diagrams,
procedures and names.
General safety warnings
The equipment is designed to operate safely when used correctly. Please
note the following safety measures to avoid injury or damage.
Equipment operation must comply with all local codes and regulations.
Original packaging should be kept for the possible return of the unit. Only
original packaging ensures optimal protection of the equipment during
transport. If it is necessary to return the equipment during the warranty
period, the manufacturer is not liable for damages caused by incorrect
packaging.
The user must immediately notify the supplier if any problem occurs during
the use of the equipment.
This product is not intended for use in areas where there is a risk of an
explosion. Do not operate the compressor in the presence of flammable
anesthetics.
Never feed oxygen or nitrous oxide into the compressor. Compressor
components are not approved for oxygen or nitrous oxide use.
Electrical system safety warnings
The equipment must be connected to ground. I order to assure proper
grounding, connect the compressor to a receptacle marked "hospital
grade."
Before the compressor is plugged in, make sure that the voltage and
frequency of the mains specified on the equipment are the same as the
power mains.
Before operating, check for possible damage to the equipment and any
connections. Damaged pneumatic and electrical lines must be replaced
immediately.
If a technical failure occurs, immediately disconnect the equipment from
the mains (pull out the main power plug).
During repairs and maintenance, ensure that:
- The main power plug is removed from the power socket
- Compressed air lines are disconnected
- All pressure has been released from the air tank
Only a qualified technician can install this equipment.

GENERAL INFORMATION
NP-DK50 DE-LF-9_02-2019 5
WARNING NOTICES AND SYMBOLS
The following symbols are used for important information in the Installation,
Operation and Maintenance Manual and on packaging and the product:
Attention, see instructions for use
Caution, risk of electric shock
Consult instructions for use
CE mark of compliance
Caution, hot surface
Handling mark on package – FRAGILE
Handling mark on package – THIS SIDE UP
Handling mark on package – KEEP DRY
Handling mark on package – TEMPERATURE LIMITATIONS
Handling mark on package – LIMITED STACKING
Mark on package – RECYCLABLE MATERIAL
Alternating current
Ground connection
Equipotentiality
Fuse

GENERAL INFORMATION
6 NP-DK50 DE-LF-9_02-2019
USE
The equipment can be installed and operated only in a dry, ventilated and
dust-free area. Climatic conditions for operation - see Technical data.
The compressor must stand on a flat and stable base.
The compressor must not be exposed to rain. The equipment must not be
used in humid or wet environments. Never use the compressor in the
presence of flammable liquids or gases.
Before connecting the compressor to respiration equipment, make sure
that it meets the requirements of the respiration equipment. Refer to the
Technical data for this purpose.
Any use other than the compressor‘s intended use is not considered to be
safe. The manufacturer is not responsible for any damages that result if
the compressor is used for any other purpose. Risk is exclusively
assumed by the operator/user.
STORAGE AND TRANSPORT
The compressor is shipped from the factory in transport packaging with the
pump stabilized, protecting it from damage during transport.
For transport, always use the original packaging and secure the compressor
in the upright position.
Protect the compressor from humidity, contamination and extreme
temperatures during transport and storage. A compressor in its original
packaging should be stored in a warm, dry and dust-free area.
Keep the packaging material, if possible. If not, dispose of the packaging
material in an environmentally-friendly way. Cardboard can be recycled.
Prior to transport it is necessary to secure the motor inside the
compressor (Chapter 5).

EQUIPMENT DESCRIPTION
NP-DK50 DE-LF-9_02-2019 7
2. EQUIPMENT DESCRIPTION
1. OUT - compressed air outlet
2. Pressure regulator
3. Output air pressure gauge (with accuracy of ± 5%)
4. Hour meter
5. Switch, power connector and fuses
6. Power cord anchor
7. Equipotential connector
8. Compressor
9. Suction filter
10. Fan
11. Safety valve
12. Filter (5 μm) with automatic condensate drain
13. Supporting base
14. Cooler
15. Regulator safety cover
The equipment uses an oil-free piston compressor (8) driven by a low-maintenance
electric motor. Compressed air is cooled in the cooler (14) where condensed water is
separated into a separate tank (12). Compressor working pressure is set using the safety
valve (11). Constant output pressure is maintained by the pressure regulator (2).

EQUIPMENT DESCRIPTION
8 NP-DK50 DE-LF-9_02-2019

TECHNICAL DATA
NP-DK50 DE-LF-9_02-2019 9
3. TECHNICAL DATA
TYPE DK50 DE
VERSION low flow
Output flow at pressure
3.5 bar (51 psig) L/min 20
Voltage / Frequency / Nominal
current V/Hz/A 230/50-60 / 2.3
110-120/60 / 4.4
Air filtration m 5
Pressure dew point at 3 bar, 20°C, 50%
Relative air humidity 5°C (9°F) below the ambient temperature
Outlet connection DISS 1160-A (3/4"-16 UNF)
Sound level dB(A) 48
49 / 60 Hz
Mode of operation Continuous - S1
Separation of condensed water Automatic
Output pressure 3.0 bar (43 psig)
Adjustable to max. 3.5 bar (51 psig)
Pressure range 5 bar ( 72.5 psig )
Adjustment of output air pressure Pressure regulator
Dimensions of compressor w x d x h 270 x 395 x 400 mm (10.5 x 15.5 x 16 in )
Dimensions of packaging w x d x h 350 x 460 x 500 mm ( 14 x 18 x 19.5 in )
Net weight 23 kg ( 51 lbs )
Gross weight 29 kg ( 64 lbs )
Implementation according to EN 60601-1, EN
12021 Class I.
Classification acc. to MDD 93/42 EEC, 2007/47
EC II b

TECHNICAL DATA
10 NP-DK50 DE-LF-9_02-2019
Climatic conditions for storage and transport
Temperature –25°C to +55°C (-13°F to +131°F), 24 hrs +70°C (+158°F)
Relative air humidity 10% to 90% (no condensation)
Relative air humidity for seaworthy packing 10% to 100% (with condensation)
Climatic conditions for operation
Temperature +15°C to +40°C (+59°F to +104°F)
Relative air humidity up to +70%
IPX0 Rating
FAD efficiency correction for differences in elevation
FAD correction table
FAD efficiency refers to conditions at an elevation of 0 mamsl:
Temperature: 20°C
Atmospheric pressure: 101325 Pa
Relative humidity: 0%
Elevation [mamsl] 0 - 1500 1501 - 2500 2501 - 3500 3501 - 4500
FAD [l/min] FAD x 1 FAD x 0.8 FAD x 0.71 FAD x 0.60
Other manuals for DK50 DE
3
Table of contents
Languages:
Other EKOM Air Compressor manuals

EKOM
EKOM DK50 F User manual

EKOM
EKOM DK50 4VR/50 User manual

EKOM
EKOM DK50 DS User manual

EKOM
EKOM DK50 Z Instruction manual

EKOM
EKOM DK50 9x4VRT/M User manual

EKOM
EKOM DK50 DE User manual

EKOM
EKOM DK50-10 Series User manual

EKOM
EKOM DK50 PLUS/M MOBILE User manual

EKOM
EKOM DK50 Plus Mobile User manual

EKOM
EKOM DK50 PLUS/M Series User manual

EKOM
EKOM DK50 2V/50 User manual

EKOM
EKOM DK50 2V/50 Instruction manual