3. UNIT INTO OPERATION
Commissioning Scaler with LED Lighting
1. Turn the water control knob (7) counterclockwise to maximum and the scaler knob (1) counterclockwise to the position (min.).
2. Connect the footswitch (11) with the socket (5) of the control unit.
3. Connect the water supply hose to the water input (8) and the other end using the supplied connector to a water source
(line, angle valve or water pressure tank) to.
4. Turn a scaler tip in the scaler handpiece and pull the top with the supplied key fi rmly
.
5. Connect the device-side power cable to the unit to the jack (6) and the power adapter to the power grid.
6. Set the desired intensity level using the adjuster (1) and start or stop the function with the footswitch.
7. Correct the desired amount of water with the adjusters (7) on the back.
Regularly check your scaler tips using the supplied picture for wear. In case of heavy wear urgently replace the tips.
Commissioning Drill/Polisher
1. Check if the power is off (the adjuster (2) and (3) to “0”).
2. Connect the Micromotor with the socket (4) of the control unit.
3. Unscrew the dust cap from the micro engine and plug the Handpiece (not included) on.
• Low Speed Dental Handpiece – Contra Angle (RA) (Item no. 174026)
• Low Speed Dental Handpiece – Straight (HP) (Item no. 174027)
4. Insert the drill bit (not included in scope of supply) in the appropriate handpiece. To lock the drills please refer to the
operating instructions of the handpieces. Check the correct position by turning the hand piece or gently pulling on the
drill.
5. Turn the mode selector switch (2) to micromotor foot operation or manual operation.
6. Use the adjuster (3) the desired direction of rotation and adjust the desired rotational speed “min. – max.” one.
7. With the footswitch (11) to start the engine or turn it off.