2.START‐UP/OPERATION
DepressthePowerbuttonontherightsideofthe
instrumenttostartoperation.Themainstartup
screenwilldisplaytoallowyoutoselectthemenu
optionoperationtoperform.Thetouch‐screen
iconoptionsaredetailedinthefollowing
paragraphs.
2.1ACQUIREICON
TheAcquireiconallowstheacquisitionofdatafroman
EbtronGoldSeriesTransmitterequippedwithanEB‐Link
optioncard.Thetransmittercovermustberemovedto
permitIRcommunicationwiththeEB‐LinkReader.When
usingtheEB‐LinkReader,remembertopointthedevice
directlyatthetransmitterwithin3.3feet(1meter)ofit
andholditinthatpositionuntiltheACQUIREeventiscomplete.
Dataacquisitiontypicallyoccursinafewseconds.
Toacquiredata:
a.PointtheEB‐LinkReaderatthetransmitteratadistanceof<3.3
feet(1meter).
b.TouchAcquirescreenicon.KeepEB‐LinkReaderpointedat
transmitter.Screenwillindicate“SearchingforGTx116”,andthen
“Eb‐LinkEstablished.AcquiringData.”Thisisfollowedby“Success.
DataAcquired.”beforesummarymeasurementdataisdisplayed.
SelectingtheTRAVscreeniconallowsindividualprobeandsensor
flowsandtemperaturestobedisplayed.
SelectingtheSAVEscreeniconallowsthecurrentdatatobesaved.
TheSAVEoptionisonlyvisibleifthedatahasnotalreadybeen
saved.
SelectingtheRE‐ACQUIREscreeniconallowsfornewdatatobe
acquiredfromthetransmitter.