Diamond OLP-82P Manual

Part No. 1902050 - Version 01.2020
FIBER OPTIC CONNECTORS
DIA-CLEAN
DIGITAL VIDEO MICROSCOPE
CLEANING PROCEDURE

Dia-Clean Digital Video Microscope - e 1
INDEX
1. GENERAL_________________________________________________________________________________2
2. HANDLING INSTRUCTIONS OF THE CONTROLL MICROSCOPE __________________________________3
3. SAFETY PRECAUTIONS TO AVOID BATTERY LEAKAGE IN THE VFL _______________________________4
4. FIBRE OPTIC CLEANING AND TROUBLESHOOTING PROCESS FLOW ____________________________5
5. CLEANING THE CONNECTOR FRONT FACE USING EASY CLEAN (DRY) ___________________________6
6. CLEANING AND INSPECTING FERRULE ENDFACE USING A ISOPROPYL ALCOHOL (WET) _________7
7 CLEANING THE CONNECTOR FRONT FACE USING “IBC CLEANER” ________________________9
8 ZONE DEFINITION ______________________________________________________________________12
9. EXAMPLES _____________________________________________________________________________13
10. CLEANING THE MATING ADAPTER SLEEVE _____________________________________________16
11. INSPECTION __________________________________________________________________________17
12. ACCESSORIES/CONSUMABLES _______________________________________________________20
0
INDEX

Dia-Clean Digital Video Microscope - e 2
1. GENERAL
Proper performance of a fibre optic connection is strongly dependant on the cleanliness of the mated ferrules.
DIAMOND connectors are thoroughly inspected and cleaned prior to shipment.
Before placing a connector in service, it is however advisable to inspect it before. After repeated matings,
however, or when degraded performance is observed, it may become necessary to clean the individual ferrules
and mating sleeve. The following instructions outline proper cleaning and inspection procedures to help ensure
optimal connector performance.
User Manual
Refer to the Viavi user manual included with the microscope.
1
GENERAL

Dia-Clean Digital Video Microscope - e 3
2. HANDLING INSTRUCTION
OLP-82P microscope specications
General technical (Typical at 25°C)
Weight: 1.2 lb (1.4 Ib for PCM version)
Dimensions OLP-82P (HxWxD): 20.83 x 152.4 x 6.35 (8.2 x 60 x 2.5 in)
Video Display: 3.5 in color LCD, 4:3 ratio
Keypad: 11+2 dome-buttons membrane panels 4 LED indicators
Connector: USB 2.0 (2 x host, Type A; 1 x device, Micro-B)
Power source: AC adaptor, battery (Alkaline or rechargeable Li ion), USB port
Run Time
Rechargeable Li ion=(minimum): 8 hours
Alkaline: 5 hours
Power mode: Active, Auto-off
Auto-shutoff time: User programmable
Charge Time
AC adaptor: 8 hours
USB port: 16 hours
Power source: USB port
Certification: CE, IEC/EN61326
Low Mag Field-of-View (FOV): Horizontal: 740 µm
Vertical: 550 µm
High Mag Field-of-View (FOC): Horizontal: 370 µm
Vertical: 275 µm
FBP probe specications
Dimensions: 140mm x 46mm x 43mm
Weight: 110g
Low Mag Field-of-View (FOV): Horizontal: 740 µm
Vertical: 550 µm
High Mag Field-of-View (FOC): Horizontal: 370 µm
Vertical: 275 µm
Live Image: 640 x 480 fps
Connector: USB 2.0 (backwards compatible to USB 1.1)
Cord length: 183 cm (6ft)
Camera sensor: 2560 x 1920, 1/2.5-in CMOS
Particle Size Detection: <1 µm
Light Source: Blue LED, 100,000 + hour life
Lighting technique: Coaxial
Power Source: USB Port
Certification: CE
Attention and Maintenance
To avoid damage, don’t disassemble and assemble the instrument or change the innercircuit and parts
Prevent liquid getting into the battery house
Protect it from violent shock
With proper care this microscope will provide years of sale, reliable service
Warranty
This product is warranted to be free of all defects in material and workmanship for a period of 12 months
from the date of delivery. The warranty does not apply to ani instrument that has become worn, defective,
damaged, or broken due to abuse, misuse, tampering, or unauthorized repairs. Under this warranty, we will
repair or replace, without charge to the purchaser, but we reserve the exclusive right to determinate the cause
of failure and advice if a warranty replacement or repair can be made. If is should become necessery to return
the instrument for service during of beyond the warranty period, the sender is responsible for shipping charges,
freight, insurance and proper packaging to prevent damage intransit.
2
HANDLING INSTRUCTION

Dia-Clean Digital Video Microscope - e 4
3. SAFETY PRECAUTIONS TO AVOID BATTERY LEAKAGE IN THE VFL
- Do not leave the VFL containing batteries in places where the temperature may rise significantly.
- When the batteries are not in use for a long period of time, be sure to remove them from the VFL and
store them in a cool place.
- Do not use batteries after the expiring date.
DIAMOND SA assumes no responsibility in case fluid acid escapes from the batteries causing the
microscope not to function any longer.
3
SAFETY PRECAUTIONS TO AVOID BATTERY LEAKAGE IN THE VFL

Dia-Clean Digital Video Microscope - e 5
4. FIBRE OPTIC CLEANING AND TROUBLESHOOTING PROCESS FLOW
Fibre Optic Cleaning and Troubleshooting
Process Flow
If after repeated cleaning the connector still does not meet the acceptance criteria, the connector must be
replaced or re-polishing.
Re-polishing should only be performed by trained personnel using Diamond polishing machines and should
only be performed on-site when absolutely necessary. Always keep the protective cap on the connector
when not in use.
4
FIBRE OPTIC CLEANING AND TROUBLESHOOTING PROCESS FLOW

Dia-Clean Digital Video Microscope - e 6
5. CLEANING THE CONNECTOR FRONT FACE USING EASY CLEAN (DRY)
Cleaning process
1. By pushing down the feeding lever a new part of the cleaning tape moves forward and the metal shutter
on the upper side of the cleaning cartridge is opened.
2. Now the connector front face is set onto the upper end of the groove and is carefully moved along on the
cleaning tape. Please observe the direction of the arrow on the cartridge.
3. Close the protective shutter by releasing the feeding lever.
4. To inspect the ferrule endface, insert the ferrule into the opening on the inspection microscope. Set the focus
so that the observed features can be compared with the example on page 10. Also refer to the samples on
page 11-12.
5. If the criteria is met, the connector may be returned to service.
6. If the criteria is not met, repeat the cleaning procedure. If after repeated cleaning the connector still does
not meet the criteria, the connector must be replaced or returned for repolishing. Repolishing should only
be performed by trained personnel and should only be performed on-site when absolutely necessary. Keep
the connectors protective cap mounted at all times when not in use.
NB: E-2000™ with spring loaded protection cap
Insert the connector in the service adapter, part no. 1019034, and procced cleaning according to the above
description.
5.1Ferrule end face dry cleaning procedure for unmated connectors
A) When the lever is pushed down, the metal protection cover automatically slides back and a new section
of tape appears, ready for use.
B) Keep the lever down, hold the ferrule front face onto the tape and slide along the entire tape in the
indicated direction .
C) Attention: the same part of the tape may be used once.
D) Close the metal protection cover by releasing the lever.
E) Inspect the connector referring to the according “Acceptance Criteria” document. If the connector
face is not clean repeat the procedure.
F) If the result is not positive even after repeated cleaning, try with the wet procedure.
5
CLEANING THE CONNECTOR FRONT FACE USING EASY CLEAN (DRY)

Dia-Clean Digital Video Microscope - e 7
6. CLEANING AND INSPECTING FERRULE ENDFACE USING A ISOPROPYL
ALCOHOL (WET)
1. Fold a precision wipe, disposable wipe three times to create a cleaning surface cushioned with eight layers
of folded material. Be careful not to touch the cleaning surface.
2. Moisten a small area of the cleaning surface with solvent, ensuring that a portion of the surface
remains dry.
3. Press the ferrule endface lightly onto the moistened portion of the cleaning surface.
4. With light pressure, rotate the ferrule and move it back and forth over the moistened surface.
5. Immediately repeat this motion on a dry portion of the cleaning surface to remove any traces of solvent.
6. To inspect the ferrule endface, insert the ferrule into the opening on the inspection microscope. Set the focus
so that the observed features can be compared with the example on page 7. Also refer to the samples on
page 8-9.
7. If the criteria is met, the connector may be returned to service.
8. If the criteria is not met, repeat the cleaning procedure. If after repeated cleaning the connector still
does not meet the criteria, the connector must be replaced or returned for repolishing. Repolishing should
only be performed by trained personnel and should only be performed on-site when absolutely necessary.
Keep the connectors protective cap mounted at all times when not in use.
NB: E-2000™ with spring loaded protection cap
Insert the connector in the service adapter, part no. 1019034, and procced cleaning according to the above
description.
6.1 Ferrule end face wet cleaning procedure for unmated connectors
A) Fold a new a lint-free, disposable wipe three times to obtain a cleaning
surface cushioned with eight layers of folded material.
B) Moisten a small area of the cleaning surface with some 99% isopropyl
alcohol, ensuring that a small area of the surface remains dry.
C) Moisten the connector with the tissue and
leave it to act briefly. Rotate the ferrule on
the moistened tissue with an axial movement
several times applying a light pressure.
6
CLEANING AND INSPECTING FERRULE ENDFACE USING A ISOPROPYL ALCOHOL

Dia-Clean Digital Video Microscope - e 8
E) Rotate the ferrule on dry zone of the tissue with an axial movement several times applying a very slight
pressure.
F) Inspect the connector referring to the according .Acceptance Criteria. document. If the connector face is
not clean repeat the procedure.
G) If the result is not positive even after repeated cleaning, we recommend re-polishing or replacing the
connector.
H) Re-polishing may be carried out only by specially trained staff and with the appropriate tools and devices.
6
CLEANING AND INSPECTING FERRULE ENDFACE USING A ISOPROPYL ALCOHOL

Dia-Clean Digital Video Microscope - e 9
The IBC™ Brand Cleaning Tools are dry cloth cleaning tools specially designed to clean single fiber connectors
residing in an adapter, faceplate or bulkhead. The dust cap of the cleaning tool acts as an adapter for cleaning
unmated connectors. The IBC™ Brand Cleaning Tools for simplex connections are simple to use and highly
effective at removing oil and dust contaminants that can negatively impact optical performance.
The IBC Brand Cleaner 1.25 mm is a device designed for cleaning the ferrule end faces of F3000 connectors
with PC and APC polishes.
Guide Cap Cover
Guide Cap Nozzle Nozzle Extender Lock
The IBC Brand Cleaner 2.5 mm is a device designed for cleaning the ferrule end faces of E-2000, SC, ST and
FC connectors with PC and APC polishes.
Guide Cap
Guide Cap Cover
Nozzle
Technique
Pre-treatment with the first click
Clean using the second click
Simple pushing motion to engage tool
Audible CLICK to alert the operator when tools is fully engaged
7. CLEANING THE CONNECTOR FRONT FACE USING “IBC CLEANER”
7
CLEANING THE CONNECTOR FRONT FACE USING „IBC CLEANER“
Table of contents
Other Diamond Microscope manuals
Popular Microscope manuals by other brands

VWR
VWR VisiScope 384 Series instruction manual

Nikon
Nikon ECLIPSE E200 POL instructions

Leica
Leica DI C800 User's manual & installation instructions

ThermoFisher Scientific
ThermoFisher Scientific Continuµm manual

ThermoFisher Scientific
ThermoFisher Scientific Continuµm manual

Olympus
Olympus SZ61 instructions