Braun Infusomat Space User manual

Infusomat® Space
Infusion Pump System
2nd Generation Software
Quick
Reference
Guide

1
NAVIGATION
Programming through screens:
The pump screens will always indicate what
keys to press.
Entering values:
Use the left and right arrow keys to move the
cursor so the desired numerical value is highlighted.
Use the up and down arrow keys to increase or decrease
the number.
Accessing Home Screen:
Pump returns to the Home screen when stopped.
Press while infusion is running to return to the
Home screen. Access all menus from the Home screen.
NOTE: Scroll Bar appears on screens when more
information is available.
CHANGE PUMP DISPLAY
1. With the pump running, press or to
choose preferred value to display on bottom left
corner of display, (i.e. drug long name, Care Unit,
volume totals, etc.)
Understanding Pump Symbols:
For a complete listing of pump symbols, refer to the last
section of this guide.
Home
Symbol Scroll
Bar

2 3
OPEN DOOR AND LOAD IV LINE
1. Press to power on.
2. Press door open key . Respond to prompt and
press to open door.
CAUTION: Keep the roller clamp closed while inserting
IV line.
3. (1) Load IV line from right to left. Insert the
two hole clip.
(2) Attach white pronged clip.
NOTE: Ensure silicone tubing segment is not twisted.
(3) Insert the green anti-free ow clip in the green
slot (with hook facing down and away from pump)
until the opening lever snaps out and the ashing
yellow light goes out indicating pump has occluded
IV line.
4. Press IV line into the air sensor guide and notches
on right and left side of the pump.
5. Close the pump door by rmly placing pressure until
you feel the pump pull the door shut.
NOTE: Do not open the roller clamp until programming
prompt appears.
CAUTION: Pump door will not close and pump alarms
if the anti-free ow clip is not properly inserted. Do not
force door closed, check IV line and anti-free ow clip
(green) for proper loading.

2 3
6. If priming is activated, press to prime on the
pump.
7. Note prompt to “Disconnect IV line from patient
before priming”, press to conrm.
8. Respond to “Priming stopped” message by conrming
with . Respond to “Repeat priming” prompt with
for “Yes” or for “No”.
PRIMARY INFUSION
1. “Continue last infusion?”, press for “Yes”or
for “No”.
2. If “Yes”, scroll to conrm each value using and
to conrm values. If no, press to program
new infusion.
3. Select Care Unit using and conrm with .
4. Select Patient Prole using and conrm
with .
5. Select “All drugs” or a drug category, using
and conrm with .
6. Select drug using and conrm with .
Navigate through the list with and or use
and to skip through the alphabet
(i.e. ABC-DEF).
7. Select drug concentration using and
conrm with .
8. If prompted, read and conrm clinical advisory
with .
9. Enter or edit infusion parameters by scrolling
to desired parameter and press to open the
value editor.

4 5
10. Use and to move to the digit to be
programmed.
NOTE: The may be used to clear existing values.
11. Use and to program new values.
12. Press to conrm value.
13. If value is outside soft limits, respond to prompt with
for “Yes” to override soft limit and for “No”.
Note change in soft limit symbol when overridden.
or .
NOTE: If value is outside hard limit, press to conrm
message and edit value.
14. Scroll down to enter and/or review VTBI and Time.
VTBI is required to start the infusion.
15. Press to begin infusion.

4 5
TITRATION
NOTE: The doserate/rate may be titrated while the pump
is running.
1. Press to open editor.
2. Program desired value using . Press to
conrm which initiates new doserate/rate.
CHANGE WEIGHT, BSA, VTBI AND TIME
NOTE: Changing VTBI or Time may change infusion rate/
dose depending on infusion mode.
1. Press to return to the Home screen if pump is
infusing.
2. Scroll to Weight, BSA, VTBI or Time using .
3. Press to select.
4. Enter new value using .
5. Conrm with .

6 7
SECONDARY INFUSION
NOTE: Prime SECondary IV line manually, close roller
clamp and connect to PRIMary IV line at needleless port
ABOVE the pump.
1. Stop PRIMary infusion by pressing .
2. Scroll to and select “SECondary” by using .
3. Four sub-menus will be displayed:
• New SECondary
• Use a Previous SEC infusion
• Back to PRIM
Determines switch back to PRIMary. “manual”
stops the pump and alarms when the SECondary is
complete. “Auto-change” converts to the PRIMary
infusion when the SECondary is complete.
NOTE: The method of switching back to PRIMary
remains for all SECondary infusions until changed.
• Basic infusion
4. Select “New SECondary” with .
5. Select drug using and conrm with .
NOTE: SECondary medications will most frequently be
set up as Dose Over Time infusions.
6. Enter Total Dose using and conrm with .
7. Enter Total Time using and conrm with .
NOTE: The pump calculates rate and VTBI.
8. Press , note message to check bag height
and unclamp SECondary. Press again to conrm
message and begin infusion.

6 7
NOTE: The SECondary bag must be hung at least 8
inches above the PRIMary, measured from the bottom
of the SECondary bag to the top of uid level in the
PRIMary.
NOTE: If SECondary rate exceeds 300 ml/hr clamp o
PRIMary to prevent concurrent ow.
9. To resume PRIMary prior to completion of SECondary,
press . Scroll to and select “resume PRIMary”
using arrow keys . Conrm message “Infusion
mode changed to PRIMary” with .
10. Press . Note message to clamp or remove
SECondary. Press again to infuse PRIMary.
NOTE: To resume SECondary, follow same steps as above.
NOTE: To clear a SECondary infusion prior to
completion, press , then Press and to
“delete SECondary”.
NOTE: Pump stores 5 of the previously administered
SECondary medications. They are cleared when the Care
Unit changes or the pump is powered o.
11. To select a previously administered SECondary,
stop PRIMary by pressing .
12. Scroll to and select “SECondary” menu using .
13. Scroll to and select “Use a previous SEC infusion”
using .
14. Scroll to and select drug using .
15. When all values are checked, press and then
press .

8 9
LOADING DOSE
1. Press to answer yes to loading dose prompt.
2. Enter loading dose amount using and conrm
with .
3. Enter loading dose time using arrow keys and
conrm with .
4. Press to access doserate/rate editor to program
continuous infusion.
5. Enter doserate or rate for continuous infusion using
and conrm with .
NOTE: “LDdos” and “Start” alternate in upper right
corner of screen.
6. Press to begin infusion.
7. The loading dose may be stopped at any time by
pressing . Press to conrm prompt with
amount delivered of total programmed.
8. Press to resume infusion, respond to prompt
to continue loading dose with for “Yes” and
for “No”.
NOTE: Once “No” is selected, continuous infusion will
begin and loading dose is no longer accessible.
BOLUS DOSE
1. Press .
2. If prompted, select “Program a bolus dose and time”
using and select with .
3. Enter bolus dose using and conrm with .
4. Enter bolus time using and conrm with .
5. Review bolus parameters and press to begin
bolus.

8 9
MANUAL BOLUS
NOTE: Drug Library limits do not apply to manual
boluses.
1. Press .
2. Select “Use Manual Bolus Feature” using and
select with .
3. Press and hold .
4. Bolus amount will count up on pump display.
5. Manual bolus is limited to 10 seconds.
6. Bolus amount delivered is displayed when 10 seconds
is complete or is released.
7. The pump returns to the continuous infusion when
the is released or 10 seconds is complete.
NOTE: The pump will automatically convert back to the
continuous infusion when the bolus is complete.
6. If bolus is stopped by pressing , press to
conrm prompt with amount delivered of total
programmed.
7. Respond to prompt to deliver remaining bolus with
for “Yes” and for “No”.
8. If pump is stopped by pressing or for red alarm,
press to conrm prompt with amount delivered
of total programmed. Restart infusion by pressing
, respond to prompt as above in step 7.
Other manuals for Infusomat Space
4
Table of contents
Other Braun Water Pump manuals
Popular Water Pump manuals by other brands

Sykes AmeriPumps
Sykes AmeriPumps GP100M Operation and maintenance instructions

DUROMAX
DUROMAX XP WX Series user manual

BRINKMANN PUMPS
BRINKMANN PUMPS SBF550 operating instructions

Franklin Electric
Franklin Electric IPS Installation & operation manual

Xylem
Xylem e-1532 Series instruction manual

Milton Roy
Milton Roy PRIMEROYAL instruction manual