Revision 4.00 / LB-0545 – CerePlex E IFU
© 2020 Blackrock Microsystems, LLC
from 0.02 Hz to 10 kHz. “Norm” (Normal) filter setting is between 0.3 Hz
to 7.5 kHz.
Reference and Ground Selector
This switch allows the user to select between different sources to use as
the input reference. If the switch is toggled to “Gnd” mode, subject ground
will be used as the input reference to the amplifier. Setting the switch
position on “Ref” (Reference), will use one of the implanted reference
wires (R1 or R2) as input reference. R1 and R2 can be selected by
another switch adjacent to reference and ground selector (see below).
R1 and R2 Reference selector
Use this switch to select which of the two available reference wires to use
as the input reference. A switch position to R1 selects reference wire
number 1 as the input reference. The R2 switch selects the reference
wire number 2 as shown in Figure 5. The Ref/Gnd switch must be in Ref
mode for this switch to be effective.
Cleaning & Maintenance Instructions
Sterilization
While not generally required, the device may be sterilized by both Vaporized
Hydrogen Peroxide and Ethylene Oxide Gas. Note that Vaporized Hydrogen
Peroxide may significantly discolor the case of the device. Newer models have a
new case that is immune to this discoloration.
Cleaning and Maintenance
The CerePlex E case may be wiped down with standard sanitizing wipes,
isopropyl alcohol (70%) or ethanol (96%). If necessary, the surfaces and
connection ports may be wiped with enzymatic detergent. Dust and debris on the
exterior connectors/ports may be removed with compressed air.
The collar assembly underneath the CerePlex E should be inspected before each
use and cleaned every time after use. To clean the collar and filament film, follow
the instructions below:
Visually inspect the connector; remove any particles if present.
Visually inspect the film. If the film is swollen or degraded, follow the
section titled “Filament Film Replacement” (see below) to replace the film.
If necessary, clean the connector with a cotton swab and 70% isopropyl
alcohol. Do not let the cleaning solution pool. Allow the solution to dry
before use.
The connector and the film must also be cleaned after use with a cotton
swab and 70% isopropyl alcohol. Do not allow the solution to pool. Allow
the solution to dry before storage.