The Model PS-2010 is a Microprocessor based Patient Simulator. It provides ECG
Simulation with four waveforms with constant QRS duration and six machine performance
testing waveforms. The following are highlights of some of the main features:
•10 UNIVERSAL PATIENT LEAD CONNECTORS
•ECG: 30, 60, 120 AND 240 BPM
•SINE: 10, 60 AND 100 Hz
•SQUARE: 0.125 AND 2.000 Hz
•TRIANGLE: 2.000 Hz
•HIGH LEVEL OUTPUT (1 V p-p)
•AMPLITUDE ACCURACY: ± 2%
•FREQUENCY ACCURACY: ± 0.5% OF SETTING
•9 VOLT BATTERY POWER
•LOW BATTERY INDICATOR
•% BATTERY LIFE INDICATOR
•OPTIONAL BATTERY ELIMINATOR
•LEAD TEST FUNCTION
•HIGH IMPACT PLASTIC CASE
•LIGHT TOUCH KEYS
OPTIONAL ACCESSORIES:
•BC20-30103 SOFT-SIDED CARRYING CASE
•BC20-21114 BATTERY ELIMINATOR (Universal)