Barnstead|Thermolyne EASYpure D7421 Guide

1
EASYpure™ RO
Reverse Osmosis System
SERIES 742
INSTALLATION,
OPERATION,
SERVICE AND REPAIR MANUAL,
AND PARTS LIST
LT742X16 • 8/20/97
Model Voltage
D7421 120
D7422 240
D7422-33 230
D7423 100 Low Pressure
D7428 120 Low Pressure
D7429 240 Low Pressure
D7429-33 230 Low Pressure
BARNSTEAD|THERMOLYNE CORPORATION

2
Figure 1 EASYpure RO
Ventgard
Back
Door
Latch Pressure
Gauge Indicator
LED's
Tank
Water
Level
Sight

3
Table of Contents
Safety Information ........................................................................................................................................4
Alert Boxes................................................................................................................................................4
Warnings....................................................................................................................................................4
Specifications................................................................................................................................................6
Feedwater Requirements1........................................................................................................................6
Product Water Specifications ....................................................................................................................6
Electrical Requirements ............................................................................................................................6
Environmental Conditions..........................................................................................................................6
Product Water Rejection Rates ....................................................................................................................6
Declaration of Conformity ............................................................................................................................7
Unpacking ....................................................................................................................................................8
Installation ....................................................................................................................................................9
Locating your EASYpure RO ....................................................................................................................9
General Usage ..........................................................................................................................................9
Bench Mounting ......................................................................................................................................10
Wall Mounting ..........................................................................................................................................11
Utility Connections ..................................................................................................................................12
Feedwater Connection ........................................................................................................................12
Reject Water Connection ....................................................................................................................13
Tank Overflow Connection ..................................................................................................................14
Connections to Equipment Supplied By the EASYpure RO. ..................................................................15
Installing The Prefilter and Membranes ..................................................................................................16
Prefilter Installation ..............................................................................................................................16
Membrane Installation..........................................................................................................................17
Electrical Connections ............................................................................................................................20
Membrane Flush......................................................................................................................................20
Set "Replace Prefilter" Indicator ............................................................................................................21
Operation....................................................................................................................................................24
Indicators ................................................................................................................................................24
Pressure Gauge ..................................................................................................................................24
LED's....................................................................................................................................................24
Maintenance ..............................................................................................................................................27
Caring For Your EASYpure RO ..............................................................................................................27
System Sanitization..............................................................................................................................27
Ventgard® Filter Replacement ............................................................................................................30
General Cleaning Instructions..............................................................................................................30
Service and Repair ....................................................................................................................................31
Normal Unit Operation ............................................................................................................................31
Fuse Replacement ..............................................................................................................................33
Membrane Replacement......................................................................................................................33
Prefilter Replacement ..........................................................................................................................34
Parts List ....................................................................................................................................................35
Recommended Spares ........................................................................................................................35
Ordering Procedures............................................................................................................................37
Troubleshooting ..........................................................................................................................................38
Wiring Diagrams ........................................................................................................................................39
Exploded View............................................................................................................................................42
Exploded View Key ....................................................................................................................................43
One Year Limited Warranty ........................................................................................................................44

4
Safety Information
Your Barnstead EASYpure RO has been designed
with function, reliability, and safety in mind. It is
your responsibility to install it in conformance with
local electrical codes. For safe operation, please
pay attention to the alert signals throughout the
manual.
Warnings
To avoid electrical shock, always:
1. Use a properly grounded electrical outlet
of correct voltage and current handling
capacity.
2. Do not mount the EASYpure RO directly
over equipment that requires electrical
service. Routine maintenance of this unit
may involve water spillage and subse-
quent electrical shock hazard if improperly
located.
3. Replace fuses with those of the same type
and rating.
4. Disconnect from the power supply prior to
maintenance and servicing.
To avoid personal injury:
Warning
Warnings alert you to a
possibility of personal injury.
Caution
Cautions alert you to a
possibility of damage to the
equipment.
Note
Notes alert you to pertinent
facts and conditions.
Alert Signals

5
SAFETY INFORMATION
1. Do not use in the presence of flammable or
combustible materials; fire or explosion may
result. This device contains components
which may ignite such materials.
2. Ensure all piping connections are tight to
avoid leakage of chemicals.
3. Ensure adequate ventilation when using
chemicals for cleaning.
4. Follow carefully the manufacturers’ safety
instructions on labels of chemical containers
and Material Safety Data Sheets (M.S.D.S.).
5. Refer servicing to qualified personnel.
C. To ensure safe mounting:
1. Wall composition and construction, as well
as fastener type, must be considered when
mounting this unit. The mounting surface
and fasteners selected must be capable of
supporting a minimum of 120 lbs.; inade-
quate support and/or fasteners may result in
damage to mounting surface and/or equip-
ment. If you are unsure of mounting surface
composition, condition and construction, or
correct fasteners, consult your building
maintenance group or contractor.

6
Specifications
Dimensions
30.5 cm W x 46.7 cm W x 46 cm H.
Feedwater Requirements1, 5
Max. Temperature ........................48.8°C
Min. Pressure (standard unit) ......30 psig
(2.1 bar)
Min. Pressure (low pressure unit) 9 psig
(0.62 bar)
Max. Pressure (standard unit)......100 psig
(6.8 bar)
Max. Pressure (low pressure unit)
15 psig (1.1 bar)
Inlet Flow Requirements ..............53.5 lph
pH ................................................4 - 11
TDS (Max. ppm as CaCO3) ........500
Silt Density Index ........................<5%
Free Chlorine................................<0.1 ppm2
Langlier Saturation Index ............Negative
Turbidity........................................<1 NTU
Iron (Total as Fe)..........................<0.5 ppm
Product Water
Specifications5
Product Water Flow Rate ............12 lph3
% Recovery..................................20%-26%
1 Feedwater suitability must be determined by a water analysis.
2 Free chlorine removed by pretreatment cartridge.
3 ±1.5 lph, feedwater at 25°C and 60 psig pressure.
4 Membrane performance is dependent on condition, pressure, recovery, water temperature and water composition. Listed membrane
performance for a new membrane is based on optimum operating conditions of 60 psig, 15% recovery, 25°C (77°F) feedwater tempera-
ture and feedwater composition of a maximum of 500 ppm NaCl at a pH of 6.0-6.5.
5 Membrane performance, flowrates and recovery information are nominal values. Variations in the feedwater temperature, pumps and
reject orifice may slightly alter these results.
Membrane Rejection Rates4
Monovalent Ions ..........................90-95%
Polyvalent Ions ............................95-99%
Particles........................................>99%
Microorganisms............................>99%
Dissolved Organics (300 MW) ....>99%
Reject Water Flow Rate 650 ml/min
at
60 PSIG ±5
PSIG
Electrical Requirements
Model D7422-33 - 230 VAC ±10%, 47-63 Hz
Model D7429-33 - 230 VAC ±10%, 47-63 Hz
Environmental Conditions
Operating: 17°C - 27°C; 20% - 80% relative
humidity, non-condensing. Installation
Category II (over-voltage) in
accordance with IEC 664. Pollution
Degree 2 in accordance with IEC
664. Altitude limit: 2,000 meters.
Storage: -25°C - 65°C; 20% - 80% relative
humidity.

7
Declaration of Conformity
Barnstead|Thermolyne hereby declares under its sole responsibility that this product con-
forms with the technical requirements of the following standards:
EMC: EN50081-1 Generic Emission Standard; EN 50082-1 Generic Immunity
Standard.
Safety: IEC 1010-1-92 Safety requirements for electrical equipment for measure-
ment, control and laboratory use; Part 1: General Requirements
per the provisions of the Electromagnetic Compatibility Directive 89/336/EEC, as amended
by 92/ 31/EEC and 93/68/EEC, and per the provisions of the Low Voltage Directive
73/23/EEC, as amended by 93/68/EEC.
The authorized representative located within the European Community is:
European manager
Barnstead|Thermolyne
Saarbrückener Str. 248
D-38116 Braunschweig
Germany
Copies of the Declaration of Conformity are available upon request.

8
Unpacking
Carefully remove the unit from its shipping contain-
er.
The following items are shipped separately and
must be removed from the packaging prior to dis-
carding the packaging:
1. EASYpure RO Ventgard air filter.
2. Power cord.
3. Pump interlock cord.
4. Membrane bypass tubing.
The following items are located inside the mem-
brane access door:
1. Feedwater tubing.
2. Reject water tubing.
3. Product tubing to connect accessory
reservoir.
4. Product tubing to connect another
EASYpure unit.

9
Locating your EASYpure RO
Your EASYpure RO must be located within six feet
of an electrical outlet appropriate for your unit, with-
in five feet of your feedwater supply and close to
an atmospherically vented drain. If you are using
the accessory 30 Liter Reservoir (Part Number
TY742X2A), the tank overflow on the EASYpure
RO's integral 6.5 liter tank should be located below
the level of the storage reservoir's overflow. This
will ensure that the EASYpure RO's internal float
switches will control the operation of the system. If
the EASYpure RO is to be used with the accesso-
ry 30 Liter Reservoir, the reservoir must not be
mounted farther than 4 feet away from the
EASYpure RO. The EASYpure RO Bench Stand
(Part Number D7424) will be required for a bench
mounted installation with an accessory 30 Liter
Reservoir.
General usage
Do not use this product for anything other than its
intended usage.
Do not mount the EASYpure
RO directly over equipment
that requires electrical ser-
vice. Routine maintenance of
this unit may involve water
spillage and subsequent
electrical shock hazard if
improperly located.
Warning
Installation

10
INSTALLATION
Bench Mounting
If you are bench mounting the EASYpure
RO without the accessory 30 Liter Reservoir
(Part Number TY742X2A), simply place the
EASYpure RO on a bench.
If you are bench mounting the EASYpure
RO and using it in conjunction with the
accessory 30 Liter Reservoir (Part Number
TY742X2A), the EASYpure RO Bench Stand
(Part Number D7424) must be added to the
EASYpure RO to ensure automatic opera-
tion. Install the EASYpure RO Bench Stand
as follows:
1. Remove the 4 feet from the
EASYpure RO and retain the
screws.
2. Place the EASYpure RO on the
Bench Stand so that the screw
holes where the feet were attached
line up with the holes in the Bench
Stand.
3. Install the 4 screws removed in
step 1 through the bottom of the
Bench Stand and into the
EASYpure RO.
4. Install the accessory 30 Liter
Reservoir according to the instruc-
tions in the accessory 30 Liter
Reservoir manual.
This manual suits for next models
6
Table of contents
Other Barnstead|Thermolyne Water Filtration System manuals
Popular Water Filtration System manuals by other brands

Wisy
Wisy LineAir 100 Installation and operating instructions

Schaffner
Schaffner Ecosine FN3446 Series User and installation manual

Pentair
Pentair FLECK 4600 SXT Installer manual

H2O International
H2O International H20-500 product manual

Renkforce
Renkforce 2306241 operating instructions

Neo-Pure
Neo-Pure TL3-A502 manual