ENGLISH | 9
MAINTENANCE
It is recommended the Ameda Planum Breast Pump be inspected at least once every year to include the following:
• Visually check pump housing for cracks, breakage, or damage and ensure Pump Feet are present.
• Visually check for power cord damage.
• Check for proper operaon of controls and buons.
• Check there are no missing screws.
• Check cover is present on Service Port Door.
• Check Tubing Adapter Port is free from obstrucon.
Quick Funconal Check
The following checks/tests should be performed prior to use by a new user:
• Visual checks for power cord damage.
• Check for abnormal noise.
• Check for proper operaon of controls and buons.
Service Port
• Do not open Service Port Door on boom of the Pump.
• Do not connect any equipment to the service port when operang the pump.
• Service port is to be accessed only by authorized Service Centers approved by Ameda, Inc.
Replacing Detachable Power Cord:
• Power cord is detachable from power inlet.
• If power cord condion is in doubt, only replace with Ameda supplied 3-conductor cord equipped with protecve
ground conductor.
TECHNICAL INFORMATION
Protecon Class II (Double Insulated), Type BF
Applied Parts
Power Supply:
Operang Input Voltage 100 VAC to 240 VAC
Operang Frequency 50/60Hz
Input Power 80VA
Fuse Rang Quick Acng (F), 2.5A, 250V, 5 x 20m
Power Cord: Use only Ameda supplied power cord
For North America, use Ameda 24502059
Minimum Voltage Rang: 125V
Minimum Current Rang: 10A
Plug Type: IEC 60320 C13 (Hospital Grade)
Maximum Length: 10 feet (3.05m)
Cord Type: 3x18AWG, rated VW-1, 105C
Jacket Type: SJT or SJTW
Cercaon: UL Listed (ELBZ)
Detachable power cord provides mains isolaon via
appliance coupler.
Operaon: Connuous
ISO 10079-1, Intermient Sucon, Medium Vacuum
Sucon: Shown in percentage of full scale sucon
1%-100%; (approximately 30mmHg to 250mmHg/
4.0kPa to 33.3kPa)
Speed: 30 to 80 cycles per minute (cpm)
Operaonal Condions: Pump
Temperature +41°F (+5°C) to +90°F (+32°C)
Relave Humidity 15% to 90%
Atmospheric Pressure 0.69atm (70kPa) to
1.05atm (106kPa)
Transport and Storage Condions: Pump
Temperature -20°F (-29°C) to +158°F (+70°C)
Relave Humidity 10% to 90%
Atmospheric Pressure 0.69atm (70kPa) to
1.05atm (106kPa)
Weight: Approximately 9.75 lbs. (approximately 4.42 kg)
Dimensions: Length 9.75 in. (24.7 cm) /Width 10.5 in.
(26.7 cm) / Height 8.0 in. (20.3 cm)
Standards:
UL Cered E493585 MEDICAL - GENERAL MEDICAL
EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY IN ACCORDANCE
WITH ANSI/AAMI ES 60601-1 (2005) + AMD1 (2012)
and CAN/CSA-C22.2 No. 60601-1 (2008) + (2014) and
IEC 60601-1-6 (2010) + AMD1 (2013) and IEC 62366
(2007) + AMD1 (2014)
-20°F
-29°C
158°F
70°C
41°F
5°C
90°F
32°C
70
106
kPa
10
90
%
15
90
%
70
106
kPa