AED HeartPlus NT-180 Guide

i
HeartPlus
™
User Manual & Warranty Registration
©2016 NANOOMTECH CO., LTD All rights are reserved. It is prohibited to reproduce or
duplicate, in any form, this manual or any part thereof without permission of NANOOMTECH
CO., LTD.
This user manual may be changed without prior notice to users for functional improvements. If
more than two (2) years have elapsed since the revision date (listed on the bottom of this
page), please contact NANOOMTECH CO., LTD or your local distributor to determine whether
additional product information updates are available.
HeartPlus™is a trademark of NANOOMTECH CO., LTD
EM0101-2 (1810-REV06)
Publisher: NANOOMTECH CO., LTD.
This issue date for the HeartPlus™User Guide, Rev. 6, is Oct 2018.
Device manufacturer
NANOOMTECH CO.,LTD.
Address: 57, Cheomdan venture so-ro, Bukgu, Gwangju, 61003, Rep. of Korea.
Authorized EU Representative
CMC Medical Devices & Drugs S.L.
C/ Horacio Lengo Nº 18, CP 29006, Málaga, Spain
Tel: +34951214054 / Fax: +34952330100 / E-Mail: info@cmcmedicaldevices.com

ii
Intentionally Blank

iii
Table of Contents
HeartPlus™-User Manual & Warranty Registration
ABSTRACT ..............................................................................................................1
SIMPLIFIED CPR....................................................................................................2
SECTION I - PRODUCT INTRODUCTION..........................................3
CONTRAINDICATIONS...........................................................................................4
PRODUCT INFORMATION.....................................................................................5
PRODUCT SPECIFICATION AND FUNCTION........................................................6
SET CONFIGURATION............................................................................................9
PRODUCT CHARACTERISTICS............................................................................ 11
DISPLAY DETAILS ................................................................................................12
NAME AND FUNCTION OF COMPONENTS..................................................... 14
SECTION II - PRODUCT USE ............................................................ 19
SIMPLE GUIDE FOR USE.....................................................................................20
HOW TO INSTALL A BATTERY PACK .................................................................21
HOW TO INSTALL ELECTRODE PADS................................................................ 23
DETAILED INSTRUCTIONS (ADULT) .................................................................. 24
DETAILED INSTRUCTIONS (PEDIATRIC)............................................................29

iv
BATTERY CHECK..................................................................................................34
PC COMMUNICATION....................................................................................... 35
SECTION III - GUIDELINES ............................................................... 37
CHECK LIST -PRIOR TO USE ............................................................................38
POST-USE PRODUCT MANAGEMENT .............................................................. 41
STORAGE GUIDELINES........................................................................................43
CAUTIONS FOR BATTERY PACK......................................................................... 44
GLOSSARY OF SYMBOLS ................................................................................... 46
SECTION IV - APPENDIX.................................................................. 49
DEFIBRILLATION WAVEFORM............................................................................ 50
CHARACTERISTIC OF ECG ANALYSIS ................................................................ 52
ELECTROMAGNETIC COMPATIBILITY (EMC)..................................................... 53
VOICE PROMPTS ................................................................................................56
DIAGNOSTICS &TROUBLESHOOTING............................................................. 57
SERVICE &WARRANTY .................................................................................... 59

1
Abstract
HeartPlus™(NT-180) is an Automated External defibrillator which restores
a normal heartbeat in patients with SCA (sudden cardiac arrest).
HeartPlus™(NT-180) is an emergency medical device.
Before using this product, please read carefully the safety precautions and
instructions for proper usage.
HeartPlus™(NT-180) is a medical device which must be tracked and
monitored to ensure optimal performance and operability.
If the HeartPlus™(NT-180) is sold, donated, lost, stolen, exported, or
destroyed, please notify NANOOMTECH CO., LTD or your local distributor.
The following visuals are used in this manual to help encourage safe and
proper usage and maintenance of the product.
Indicates possibility of a potentially dangerous
situation resulting in death or severe injury in the
case where instructions are not followed.
Indicates possibility of injury or product damage
that results in the breach of warranty agreement.
Indicates current manufacturer policy for proper
use of the product and accessories.

2
Simplified CPR
OHCA Chains of Survival
1
The recommended compression rate is at least 100 compressions per
minute.
Be sure to achieve the full compression depth of at least 5cm
(2 inches) for adult patients.
For children, be sure to achieve the full compression depth, which
should be at least 1/3 of chest anteroposterior diameter of
approximately 4cm (1.5inch) for infants and approx. 5cm (2inch) for the
child.
Achieve enough compression speed and depth, allow the chest to
recoil completely after each compression, minimize interruptions in
chest compression, and refrain from excessive artificial respiration.
Perform the chest compression before one-man rescuer tries artificial
respiration. (Perform in order of chest compression 30 times, then
followed by artificial respiration 2 times.)
1
Source: 2015 America Heart Association (AHA) guideline for CPR and cardiovascular first aid.

3
Section I
Product Introduction

Section I -Product Introduction
4
Product Introduction
Product Description
HeartPlus™is an AED (Automated External Defibrillator) which provides
defibrillation treatment for quick resuscitation of an SCA (Sudden Cardiac
Arrest) patient. This product is a portable one operating on batteries. It is
reasonably light and can be operated even by the user who receives
minimal or no education/training on the use of the product through voice
commands that guides the user through the operation sequence.
SCA (Sudden Cardiac Arrest) is the cessation of normal circulation of the
blood due to the failure of the heart. SCA can occur to anyone at anytime
and anywhere. SCA patients do not show any warning signs or symptoms.
Some individuals are at a higher risk of suffering from SCA than others
due to hereditary or other reasons.
When to Use
Patient does not respond to voice.
Patient has no reaction even when shaken.
Patient does not breathe normally.
If you suspect that the patient shows the above symptoms, follow the
voice commands after turning on this product at every stage (attaching
electrode pad, shock or no shock, CPR).
Contraindications
None known.
Product User
HeartPlus™should be used by a person who has been educated on the
use of AED (Automated External Defibrillator) and CPR (Cardiopulmonary
Resuscitation).

Section I- Product Introduction
5
Product Information
Product Name: AED
Brand Name: HeartPlus™
Model Name: NT-180, NT-180Y, NT-180B, NT-180G
Product Appearance:
Size: 293mm(W) X 291mm(L) X 71mm(H)
Weight: 1.9kg (Incl. cartridge)

Section I -Product Introduction
6
Product Specification and Function
Defibrillator
Discharge Current Waveform: BTE (Biphasic Truncated
Exponential)
Discharge Time: Automatically adjusted energy for
patient impedance
Output Energy:
-180J (for adult, load impedance of 50)
-50J (for infant, load impedance of 50)
Patient Insulation: BF Type
Charging period: Less than 13 seconds after pads
attachment.
Operation Mode: AED (automated external
defibrillator), which can analyze ventricular fibrillation
Change Method to Pediatric Mode: Power on device
in Pediatric Mode (small yellow button)
ECG
Induction Method of ECG: LEADⅡ
All-in-one Type
Cartridge
(Pre-connected
Battery And
Electrode Pads)
Basic
Specification
Composition: Disposable Electrode
pads, Battery
Frequency of Use: One time
Packing Method: Sealing Packaging
Size: 150(W) X 110(L) X 33(H)mm
Warranty: 2 years (Sealed storage at
room temperature)
Electrode
Pads
Frequency of Use : Single Use
Size: 125mm X 95mm
Area: 76.63㎠(PE Foam, Hydrogel)
Wire Length: 1.2m
Attachment point: Showed in picture on
Table of contents