PARAMOUNT BED PA-90000 Series User manual

7A10625900A4
PARAMOUNT BED CO, LTD.
●
●
●
●
●
●
INSTRUCTION MANUAL
PA-90000 Series Bed
Preface
The correct operang procedures should be explained not only to
the persons using the bed but also to caregivers.
Aer reading this instrucon manual, store it where it can be easily
accessed for later reference.
Users and/or carers are requested to report informaon about any serious
incident related to this bed that occurs within the EU to the manufacturer
and the regulang authority of the member country of residence.‘Serious
incident’ means any incident that directly or indirectly led, might have led
or might lead to any of the following:
a) The death of a user, carer or other person.
b) The temporary or permanent serious deterioraon of a user’s, carer’s or
other person’s state of health.
c) A serious public health threat.
The actual product purchased may differ from detailed descripons
given in this instrucon manual due to product improvements.
Value in the specificaon such as working range, dimensions, angles
and mass are subject to normal manufacturing tolerances. Besides some
exempons, display purpose words such as “approximately”, “about”
etc. are omied.
For any queries, please feel free to contact your dealer directly.
Before using this bed, please read this instrucon manual to fully
understand the safe and proper operang procedures.
Thank you for purchasing the PA-90000 Series Bed.
This instrucon manual describes precauons and how to safely use
and operate the bed.
●

CONTENTS
1INTENDED USE / PART NAMES
CONNECTING TO THE POWER SUPPLY (ELECTRIC HOSPITAL BED ONLY)
OPERATION OVERVIEW
USING THE HANDHELD CONTROLLER (ELECTRIC HOSPITAL BED ONLY)
OPERATION PROCEDURES
ASSEMBLING THE BED (PA-90000)
TROUBLESHOOTING
REGULAR INSPECTION AND DAILY MAINTENANCE
2
3
6
Page
1
7
OPTIONAL ACCESSORIES
8
10
11
LONG-TERM STORAGE
MATTRESS (OPTIONAL)
10
SAFETY PRECAUTIONS
SPECIFICATIONS
INSTALLING THE BED
4
5
12
13
14
15
16
1
AFTER-SALES SERVICE
2. IV pole
4-12
13-16
21
22
22
26-27
28
30
31
32
33
38
40
41-43
44-45
46
47
17
24
24
25
29
34
35-36
PRODUCT SYMBOL DEFINITIONS
39
18
4. Oxygen cylinder holder 37
3. Baery (Electric hospital bed only)
BED SIDE TABLE
OVER BED TABLE
19
1. Recommended accessories table
9
1. Aaching/detaching the head board and foot board
2. Aaching the maress stopper
1. Fing component for assembly
1. Power outlet
2. Using the cable storage hooks
1. Handle operaon (Manual hospital bed only)
2. Adjusng the angle of leg secon base
3. Locking the casters
4. Emergency CPR (Back-lowering) funcon (Electric hospital bed only)
2-3
23
ASSEMBLING THE BED (PA-90000VIP)
2. Head board, foot board, side panel installaon 23

INTENDED USE / PART NAMES
1. INTENDED USE
2. PART NAMES
The page number shown next to this
symbol indicates where the related
2
1
Handheld Controller
Head Board
Cable Storage Hooks
Caster
Caster Switching Pedal
Bed with The Central Locking System
Bed with Caster with a stopper
(When the foot board is off)
3-Crank Manual Hospital Bed
(When the foot board is off)
Main Frame
Foot Board
Handle
Caster With Stopper
Power Cable
p.26-27
p.22
p.24
p.22
p.29
p.30
p.28
p.22
p.30
(Applied part)
Part of ME EQUIPMENT that in
normal use necessarily comes
into physical contact with the
※Please refer to the「SPECIFICATION] for the paent target groups and the intended users.
【PA-90000】

The page number shown next to this
symbol indicates where the related
3
p.23
【PA-90000VIP】
Head board
Handheld controller
Leg secon base
Side panel
Caster
Matress stopper
Foot board
Hip secon base
Knee secon base
Power plug
Back secon base
p.26-27
p.29
p.22
Caster switching pedal
p.30
p.23
p.23

Be sure to read these safety precauons beforehand to ensure proper usage.
The safety precauons described here are intended to ensure safe use of the product and prevent
personal injury.
Precauons are classified as “WARNING”, “CAUTION” to indicate the degree of hazard or injury that may
result from improper use. Both are important safety precauons that must be strictly observed. Details
of the pictograms used are shown below. Be sure to fully understand these precauons before reading
the instrucon manual.
Indicates acons that may result in death or serious injury
(broken bones, pressure or paralysis) if the indicaon is ignored.
WARNING
CAUTION
Indicates acons that may result in light personal injury (bruises,
wounds, or cuts) or property damage if the indicaon is ignored.
※Aer reading this instrucon manual, store it in a place that can be easily accessed for later reference.
22 SAFETY PRECAUTIONS
4
[Example pictograms]
This symbol indicates warning or cauonary details. Specifics details are
indicated inside the symbol. (The symbol on the le indicates “Cauon: Electric
shock.”
This symbol indicates acons that are prohibited. Specific setails are indicated
inside or below the symbol. (The symbol on the le indicates “Do not
disassemble.”)
This symbol indicates instrucons or acons that are mandatory. Specific details
are indicated inside the symbol. (The symbol on the le indicates “Unplug
the power plug.”)
[Safety label]
Precauons requiring special aenon by persons
using the bed are affixed to the foot-end frame.
Do not remove or dedace the label.
If the safety label is removed or defaced, replace
with a new one obtained from your dealer.
Electric shock
Do not dissasemble

WARNING
2
5
2SAFETY PRECAUTIONS
Do not put any body part (especially the head or neck) between a fold-down side rail or board and the
raised base ( ).
Even when side rails are used, pay aenon to prevent paents from falling of the bed
A paent may fall off the bed through the gap between a board and fold-down side rail and get
injured.
If a paent bends forward over a fold-down side rail, they may fall off the bed and get injured.
Do not extend the head, hands, or feet outside of the bed.
Do not put any body part (especially the head or neck) between a fold-down side rails and the base or maress.
When the head or neck enters the gap, the paent might not be able to withdraw, and there is a risk
of injury.
A hand or finger may get caught in the gap, resulng in an injury.
Do not insert a head (fingers) and between the bed’s frame and secon base during powered operaon.
Lock the casters at all mes except when moving the bed.
When the head or neck enters the gap, the paent
and/or user might not able to withdraw, and there is a
risk of injury.
Due aenon should be paid to paents who cannot
maintain posture by themselves.
If the bed is operated with a body part (especially
head or neck) placed in the gap, the part may get
caught and injured
The body part may get caught between the lowered
base and the bed’s frame or board and get injured.
The paent may be injured geng on or off the bed
should it move.
Unlock the casters when moving the bed.
Before a paent gets on or off the bed, push down the safety stoppers of the side rail and completely
fold down the side rail.
The paent leg’s could be caught by the side rail, causing
the paent to fall off the bed or the paent’s fingers
could be caught between the supporve bars, causing
injury.
The tubes could be squashed.
Releasing the hands in the middle of the operaon or
slamming the rail down could cause injury or damage
to side rail.
When using IV hanger rod, pay aenon to prevent tubes from geng caught by the side rail.
When folding the side rails, proceed slowly.

WARNING
SAFETY PRECAUTIONS
2
Do not use with infants.
They may fall down through the space of a fold-down
side rail.
Do not allow dust to accumulate on the power plug.
When dust adheres to the surface of the power plug,
the moisture it holds will allow a current to flow, there
will be a worsening of the insulaon condion, and a
risk of fire.
Wipe away any dust accumulang on the power plug
pins and mang surfaces using a dry cloth.
Do not damage cables (including the power cable).
Damage to cables may cause electric shock or a fire.
Prevent cables from geng caught by the bed’s
movable parts.
Do not put heavy objects on cables or apply strong
force
Do not place the bed on cables.
Request repair (or replacement) for damaged cables.
Do not trip over the power cable or cables of the handheld controller.
Damaged plug or cable may induce electric leakage, and in turn, electric shock.
There is a risk of falling down and geng injured.
Do not run cables (including the power cable) of the bed or other electric appliances under the bed.
The cables may get caught in a caster or movable part and damaged, causing a fire or electric shock.
When adjusng the base angle or the bed height, pay aenon to the paent’s condions.
If a paent on the bed moves when adjusng the base angle, they may fall from the bed and get
injured.
Perform the adjusment while supporng the paent who cannot maintain posture by themselves.
If you cannot watch the paent, flaen the base angle and lower the bed to the lowest posion.
If not, the paent may fall off the bed or get caught in a space of a fold-down side rail or board and
get injured.
6
Electric shock

WARNING
Do not allow operaon of the bed by a child aged 12 or under, or someone deemed incapable of
comprehending the operaon.
When transporng a paent, aach the fold-down side rails.
Do not perform an angle adjusment when the paent is lying face down.
Do not stand up on the bed.
Do not go under the bed or put your head, hand, or foot under the bed.
7
When a child aged 12 or under, or somenone deemed incapable of comprehending the operaon
(such as someone having a cognive illness) operates the handheld controller by themselves, there
is a risk of unexpected injury such as the body becoming caught in the bed by mistake.
If there is a risk of such person touching the handheld controller, disconnect the power plug in order
to prevent an accident caused by incorrect operaon.
When a child aged 12 or under, or someone deemed incapable of comprehending the operaton
(such as someone having a cognive illness) operates the handheld controller by themselves, there
is a risk of unexpected injury caused by geng caught in the bed by mistake. (3-crank gatch bed
only)
There is a risk that the paent could fall from the bed and become injured.
Do not move the bed with holding the fold-down side rails. The side rails may get damaged or
deformed.
Use the bed in the proper orientaon.
Lying down on the bed with incorrect orientaon of
head end and foot end will result in an unnatural
posture when the bed posion is adjusted and this
may cause injury.
Bending the joints in the wrong direcon may result
in injury.
There is a risk of injury due to falling or toppling from
the bed.
You may get caught between the bed’s movable part
and the frame or fold-down side rail, resulng in
injury. Prior to operang the bed, make sure that
there is no obstacles around the bed.

WARNING
SAFETY PRECAUTIONS
2
Do not insert the power plug with wet hands.
8
There is a risk of electric shock due to short-circuing.
Do not spill liquids on the bed.
Do not spill liquids on the bed’s electrical components, such as the motors, handheld controller, or
nurse control panel, otherwise, there is a risk of malfuncon. If liquids are accidentally spilt, unplug
the power plug from the power outlet and contact your dealer.
Do not use a mobile phone or other wireless device in the vicinity of the bed.
This may cause a malfuncon of the bed.
Repair or modificaon is not to be performed by the customer.
There is a risk of abnormal operaon and resulng
injury.
Always unplug the power plug when using electronic medical equipment.
Do not aach a restraint belt to the main frame or side rails.
Operaon in posion such as with the back raised or the knee raised may squeeze the body and
cause injury.
Use the bed properly according to the paent’s medical condion.
Improper use of the bed may adversely affect the paent’s medical condion.
Before the paent or a family member operates the bed, make sure that the doctor or nurses explain
related details given in the instrucon manual and proper usage to suit the paent’s medical
condion.
When leaving the bed, lower the bed to the minimum height.
If a doctor, nurse, or aendant leaves the bed, lower the bed as much as possible according to the
paent’s condions because there is a slight chance that the paent may fall off the bed.
Using electronic equipment (e.g. microwave or
high-frequency medical equipment) at the same me
may cause the bed to malfuncon.
When the bed will be used in conjucon with other
medical equipment, check the safety consideraons
before use.
Electric shock
Do not dissasemble
Unplug

9
the wall, beam, or power outlet on the wall.
CAUTION
When using the bed, make sure that the beds does
not hit the wall beam, or power plug during the
Make sure that there is at least 20 cm clearances
from the wall or beam.
bed could be damaged or deformed and result in
injury.
bed could be damaged or deformed and result in
injury.
The bed should not be used by two or more persons.
This bed is designed for use by one person.
be damaged and result in an injury.
The maximum weight the bed can bear is 204 kg
If a doctor or nurse needs to get on the bed for a
beforehand.
The load to be applied to the bed is less than the
to have the bed periodically inspected
Secure at least 20cm clearance
Other manuals for PA-90000 Series
1
Table of contents
Other PARAMOUNT BED Indoor Furnishing manuals

PARAMOUNT BED
PARAMOUNT BED L'za Series User manual

PARAMOUNT BED
PARAMOUNT BED cococia Riraku flow User manual

PARAMOUNT BED
PARAMOUNT BED PS-0232S User manual

PARAMOUNT BED
PARAMOUNT BED PE-6001 User manual

PARAMOUNT BED
PARAMOUNT BED PF-1100A Series User manual

PARAMOUNT BED
PARAMOUNT BED PM-30 Series User manual