NimbusTM EU Patient Manual --560-IFU-IT1098, Rev. A 9
Warnings, continued
Always confirm that the established infusion route matches the
intended administration route prior to the start of the infusion.
Clearly differentiate established administration routes when using the
pump with a patient that has multiple administration routes established.
Do not operate the pump outside of the environmental limits detailed in
the pump specifications.
Upon starting an infusion, the slide clamp should be opened.
Before packaging the drug/fluid bag and pump into the pouch, check
the fluid path for kinks, closed clamp or other occlusions.
Do not put excessive pressure on the pump that may cause the device,
bag or tubing to compress (i.e. leaning against pouch, sleeping on
pouch, etc.).
Do not leave the pump or cassette/administration set unattended at
home or other places where children or pets may come in contact with
the device.
Disconnect the administration set from the patient before removing the
cassette from the pump.
Always close the slide clamp before removing the cassette from the
pump.
Do not store the pump with a cassette/administration set attached.
Do not place objects over the pump that may muffle the pump.
If the pump is not functioning as expected, the pump should not be
used, and should be returned for inspection or repair.
There are no user serviceable parts inside the pump. Refer all services,
repairs, and calibrations to qualified technical personnel. Do not make
unauthorized modifications to the infusion pump.
Avoid spills on the pump. Follow the instructions specified in Section 6
of this manual for cleaning and disinfection of the pump.
To avoid mechanical or electronic damage, do not steam, autoclave, or
immerse the pump in any fluids or cleaning solutions, and do not spray
such fluids directly on the pump.
Failure to properly follow the cleaning instructions may result in an
electrical hazard, damage to the pump, and voided warranty coverage.
Do not use this device for neonates.
If the pump is dropped or hit, inspect the pump for damage.
Immediately stop using the pump if it is physically damaged or not
functioning properly. Contact distributor to return the pump.
Avoid using and/or storing the pump where it may come in contact with
excessive lint, dust, direct sunlight, heat and moisture.