ENGLISH
11
INSTALLATION
CE conductivity measurement range and K
factor (cell constant)
▪ K1 Range up to 20.00 mS
Hysteresis / PWM start-up point = 500 μS
Resolution 10 μS
▪ K5 Range up to 2.,000 μS / Hysteresis / PWM start-up point = 50 μS
Resolution 1 μS
▪ K5 Range up to 200 μS / Hysteresis / PWM start-up point = 0.50 μS
Resolution 0.1 μS
Temperature settings: Manual or automatic offset (auto with temperature probe PT100)
▪ Resolution 0.1% °C ▪ Precision: ± 0.5% °C
Temperature probe range: – 20.... 100°C
Power supply / Consumption: Universal power supply 100250VAC / 5W at 240VAC
Microprocessor technology: SMD components with a 6-key digital control keypad
Linearity, Stability, Reproducibility: 0.5% in standard conditions
Display: Backlit 126x64 display; Visible display area 70x37 mm
Delay on Setpoint: Relay activation delay, programmable for each setpoint (999 sec.)
Delay on start-up: Delay in relay when the unit is switched on, programmable
Consumption / Rated Current: 230Vca 5W = 25mA ▪ 24Vca-cc=5W = 230mA ▪ 12Vcc 5W = 460mA
Internal electrical protection: Power supply unit assures electrical protection (instead of fuse)
Level / Relay remote control: Chemical additive level (level probe not included) output voltage +5VDC
RELAY 1 output: SV (solenoid valve) PURGE ON-OFF / voltage-free, relay 5Amax 230VAC
RELAY 2 output: INHIBITOR ON-OFF metering pump/ voltage-free, relay 5Amax 230VAC
RELAY 3 output: BIOCIDE1 ON-OFF metering pump/ voltage-free, relay 5Amax 230VAC
RELAY 5 output: BIOCIDE2 ON-OFF metering pump/ voltage-free, relay 5Amax 230VAC
AUX Output: ALARM voltage-free contact, relay 5A max 230Vac
FLOW sensor: Blocks outlet operations if there is no ow in the probe socket.
Outputs: Output 0/4...20 mA1 EC: Adjustable (500 maximum input impedance), with galvanic separation.
Connected to Conductivity measurement settings.
Output 0/4...20 mA2 EC: Adjustable (500 maximum input impedance), with galvanic separation.
Connected to Conductivity measurement settings.
Load: Resistive load 5A at 230VAC / Inductive load 0.5A at 230VAC
Relay insulation voltage: > 3000VAC
Contact relay duration: 5x104operations (5A at 230VAC)
Noise level: Irrelevant
Operating temperature: ideal temperature 5°C-40°C, resistance up to 0°C-45°C
Environmental conditions: Possibly dry environment, altitude up to 2000m, Relative humidity 80% for temperature up to
31°C linearly decreasing to 50% of relative humidity at
40°C. Pollution degree 2.
Transport / storage conditions: – 560°C in a dry environment
Install the instrument in a dry place, away from heat sources at
a maximum room temperature of 40°C.
Comply with standards in force in the different countries
regarding electrical installation (Fig. 2). If the power cord does not
have a plug, the device must be connected to the mains by
means of an omnipolar disconnecting switch with at least 3
mm between the contacts. All the power supply circuits must
be interrupted before accessing the connection devices.
Fig. 1
–
Electrical connection
TECHNICAL SPECIFICATIONS OF THE INSTRUMENT